P04 INSTRUMENT HISTORY
 
PRP   INSTRUMENTS    SOAR   SOAROPS
 
2000-03
System shipped RSMAS to BNL for preparation for calibration.
Note: condition from RSMAS was very bad. It had considerable soot/oil contamination. The FRSR diffuser was very dirty. We cleaned it as much as we could, but it still seems a little off color.
This unit has an old 848 B&W pyranometer. It appears that water and salt have entered the instrument. One can see salt crystals inside the dome. On inspection, the o-ring for the desiccant plug seems to have been broken and repaired by adding another o-ring on top of the old one. This does not seal properly.
2001-04-29
Electronic re-calibration. Prpcal_04_0104 is almost identical to PrpCal_04_med00 done a year ago.
2001-05-04
Deployed on Mt. Lemmon for calibration.
2001-06-12
Returned to BNL. Heater circuit problems.
2001-06-20
At BNL, heater problem resolved.. Corrosion in the cable was found right at pin 1 (the head thermistor.)
2001-06-21
Ship to RSMAS for Gulf of Mexico Cruise on R/V Justo Sierra. Expected return 1 Aug.
2001-11-27
Changed shaft o-rings and the motor case cover. The cover now has a "D" hole for the Conxall.
2002-01-02
Examined the erroneous head temperature readings. The first head temp. reading taken by the A to D after the system goes into LOW MODE is bad while the rest are normal. We've decided to leave this alone and continue to monitor the readings.
2002-01-16
Re-calibrated the electronics in preparation of taking the unit to the Akademic Ioffe in Ushoya. Installed on the roof.
2002-05-28
Received P04 from RSMAS today for calibration and maintenance.
2002-08-07
Sent head 462 to YES for calibration. Received a new PSP from RAMAS to replace the 8-48 sensor that was there. Did a thermistor calibration on the PIR. Roof tested for about 2 weeks. Shipped to Bob Evans on Thursday 08/29/02.
2003-04-27
Received at BNL for maintenance.